Envirotronics Star Galaxy 44 is a test chamber with an integrated vibration table that has been specially designed to perform HALT and HASS processes. Star Galaxy chambers are equipped with a Liquid Nitrogen ( LN2 ) cooling system. The LN2 cooling system, with modulating valve and directed airflow to the product, provides the rapid thermal change rates required to achieve maximum product stress. Additionally, these rates are accomplished with smaller space requirements, lower audible noise, no water requirement, and lower maintenance costs than a typical refrigeration system.

The large mounting area of the 44 in. table, comprised of a product mounting configuration of 3/8 in. holes in a 4 in. by 4 in. mounting pattern, allows great flexibility in mounting your product or fixturing. Our improved-efficiency pneumatic vibrators deliver your required vibration energy while utilizing 20% less air. This translates into savings for you.
The combination of Envirotronics’ control system, which gives the user the flexibility to control via one accelerometer or any combination of 4 accelerometers, and our adjustable air flow package that allows the user to redirect airflow to the product under test puts control in the hands of the Star Galaxy user. It’s time to shoot for something better -
Shoot for the Stars!
»Test Chamber
- Workspace dimensions:
- W 54 in x D 54 in. x H 58 in.
- W 1371 mm x D 1371 mm x H 152.4 cm
Note: Chamber dimesions on international models may vary.
- Temperature range: -100˚C to +200˚C
- Transition rate: 70˚C/min.
- LN2 System: Low pressure (Press. 50 Psig)
LN2 cooling system with proportionately actuated control valve
- Adjustable airflow package to allow user to redirect airflow to product under test
- Both front and rear full opening doors
- Four (4) 20 in. x 20 in. multipane viewing windows, two in each door
- Two (2) interior lights
- ETL (UL STD 508A) approved main power panel
- One (1) 6 in. x 12 in. access port with insulating plug (centered in left hand side wall)
»Vibration Table
- Triaxial, pseudo-random, 6 DoF vibration
- Vibration energy: 3 gRMS to 30 gRMS
- Maximum Vibration Levels:
(Multiaxial average - X, Y, Z)
- 100 lb. Load: 30 gRMS
- 250 lb. Load: 20 gRMS
- 375 lb. Load: 15 gRMS
- 500 lb. Load: 10 gRMS
Note: Acceleration levels in excess of 50 gRMS can be achieved in the Z-axis under certain conditions.
- Minimum Vibration Levels:
(Multiaxial average - X, Y, Z)
- 100 lb. Load: 3 gRMS
- 250 lb. Load: 2 gRMS
- Vibration frequency: 20 Hz to 10 kHz
- Vibration Uniformity:
- X-axis: Standard deviation < 0.7gRMS
- Y-axis: Standard deviation < 0.7gRMS
- Z-axis: Standard deviation < 1.2gRMS
- X,Y,Z average; Std. dev. < 0.5 gRMS
(measured on an empty table at 10 gRMS)
- Noise Level: 78 dBa at 25 gRMS
- 16 pneumatic hammers (vibrators)
- Anodized aluminum mounting surface
- Mounting surface: 44 in. x 44 in.
- 3/8 in. mounting holes on 4 in. x 4 in. centers
- Instrumentation Air: 95-100 Psig, 5 CFM
- Operating Air: 110 Psig, 150 SCFM
- Maximum load capacity: 500 lbs. / 227.3 Kg
»Control
- PC, Monitor, Keyboard and mouse
- ›StarVIEW control software
- Two (2) four-channel input module for IOtech input system:
- One (1) single-axis accelerometer with mounting block, for control of the vibration table.
(operational temperature range for accelerometer(s) is -51.1˚C to +107˚C)
- DBK-18 four-channel low pass filter card
- PCB signal conditioner
- DBK-10 card cage and CA37-2 cable
- Integration into the data acquisition system
- One running-time meter for vibration run-time monitoring
Please direct any inquiries to:
›sales@envirotronics.com
*specifications for export models may vary.
|

»Standard Features of the Star Galaxy 44 HALT/HASS System include:
- 44" x 44" random vibration table with 16 hammers
- Two full opening doors (one front - left hand hinged, one rear - right hand hinged)
- Four 20" x 20" windows - two front and two rear
- One 6" x 12" access port with plug and phenolic cover
- LN2 Cooling with proportionately actuating control valve
- LN2 inlet vacuum jacketed
- Interior lights
- ETL (UL STD 508A) approved main power panel
- Chamber-mounted PC with StarView control Software
- StarView software and control package with
four-input spectrum analyzer for vibration data, including one accelerometer for control of the vibration table
(operational temperature range for accelerometer(s) is -51.1˚C to +107˚C)
- One running time meter for vibration monitoring
- Onsite installation and startup**
**Onsite installation and startup is available for our domestic customers only.

Galaxy Air Manifold
For information about our Star Plus Series go to:
Star
Plus System ›StarVIEW
Control Software
Information
Star
Plus System ›Chamber and
Console
Information
Star Plus ›Pneumatic Vibration Tables
Star Plus System ›Intro Page
Here are More Solutions from our Partner, Intertek ETL SEMKO
Learn about the new Failure Mode Verification Testing (FMVT®) Process which can produce reliable test results in as little as one day.
Get information about the ›FMVT® Machine - a patented, 6-axis, pneumatically actuated vibration machine with a uniform random frequency range of 2-5000 Hz and a total displacement of 4 inches - ›Click here.
To link to Intertek ETL SEMKO's home page
click the blue box.

|